Summary for Proteome Profiler Human NFkB Pathway Array
Kit Summary
A membrane-based antibody array for the parallel determination of the relative levels of selected human NFkB Pathway proteins. Validated for analyte detection in cell lysates.
Detects 41 human proteins and 4 serine or tyrosine phosphorylation sites simultaneously
Requires no specialized equipment
Principle of the Assay
Carefully selected capture antibodies have been spotted in duplicate on nitrocellulose membranes. Cell lysates are diluted and mixed with a cocktail of biotinylated detection antibodies. The sample/antibody mixture is then incubated with the array. Any protein/detection antibody complex present is bound by its cognate immobilized capture antibody on the membrane. Streptavidin-Horseradish Peroxidase and chemiluminescent detection reagents are added, and a signal is produced in proportion to the amount of protein bound. Chemiluminescence is detected in the same manner as a Western blot.
Why Use an Antibody Array to Detect Multiple Proteins?
Determining the expression of multiple Proteins in a single sample can be expensive, time consuming and can require specialized equipment. Performing multiple immunoprecipitations and Western blots requires time, labor, and reagents. The use of a multiplex antibody array to detect multiple proteins in a single sample can be cost-effective and also save time and sample.
Kit Contents
4 Array Membranes
4-Well Multi-dish
Array Buffers
Wash Buffer
Detection Antibody Cocktail, Human NF kappa B Pathway Array
Streptavidin-HRP
Chemiluminescent Detection Reagents
Transparency Overlay Template
Detailed Protocol
Stability and Storage
Reagents are stable for 12 months from date of receipt when stored in the dark at 2° C to 8° C.
Simultaneously detect the levels of these proteins in a single sample.
Product Specifications for Proteome Profiler Human NFkB Pathway Array
Species
Human
Source
N/A
Scientific Data Examples for Proteome Profiler Human NFkB Pathway Array
Detection of Human NFkB Pathway Proteins in HCT-116 Cell Line.
HCT-116 human colorectal carcinoma cells were untreated or treated with 1 µM Epoxomicin (Boston Biochem, Catalog # I-110) for 6 hours (500 µg lysate, 5 minute exposure). RS = Reference Spots.
Detection of Human NFkB Pathway Proteins in Daudi Cell Line.
Daudi human Burkitt’s lymphoma cells were untreated or treated with 500 U/mL IFN-alpha 2A (R&D Systems, Catalog # 11100-1) for five minutes (1 mg lysate, 10 minute exposure).
Detection of Human NFkB Pathway Proteins in HT-29 Cell Line.
HT-29 human colorectal carcinoma cells were untreated or treated with 50 ng/mL recombinant human TNF-alpha (R&D Systems, Catalog # 210-TA) for ten minutes with or without pretreatment with 5 µM MLN4924 (Boston Biochem, Catalog # I-502) for 1 hour (400 µg lysate, 5 minute exposure).
Preparation & Storage
Shipping Conditions
The product is shipped with polar packs. Upon receipt, store it immediately at the temperature recommended below.
Storage
Store the unopened product at 2 - 8 °C. Do not use past expiration date.