Skip to main content

Estradiol Parameter Assay Kit Best Seller

R&D Systems, part of Bio-Techne | Catalog # KGE014

R&D Systems, part of Bio-Techne
Catalog #
Availability
Size / Price
Qty
Loading...
KGE014

Control Products Available

Key Product Details

Assay Length

3.5 hours

Sample Type & Volume Required Per Well

Cell Culture Supernates (100 µL), Serum (100 µL), EDTA Plasma (100 µL), Heparin Plasma (100 µL)

Sensitivity

12.1 pg/mL

Assay Range

12.3-3000 pg/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma)

Product Summary for Estradiol Parameter Assay Kit

The Parameter Estradiol Immunoassay is a 3.5 hour competitive enzyme immunoassay designed to measure Estradiol in cell culture supernates, serum, and plasma.

Product Specifications

Format

96-well strip plate

Measurement

Quantitative ELISA

Detection Method

Colorimetric - 450nm (TMB)

Conjugate

Biotin

Reactivity

Multi-Species

Specificity

Measures Estradiol levels in various samples.  This assay cross-reacts 0.07% with 17 alpha-ethynylestradiol, 0.03% with 17 beta-estradiol 3-Benzoate, 0.86% with Estriol, 0.26% with Estrone and 0.06% with Progesterone.

Cross-reactivity

< 0.5% cross-reactivity observed with available related molecules. Cross-species reactivity observed with 1 or more species tested.

Interference

Interference observed with 1 or more available related molecules.

Precision

Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision.

Inter-Assay Precision (Precision between assays) Three samples of a known concentration were tested in twenty separate assays to assess inter-assay precision. Assays were performed by at least three technicians using two lots of kit components.

Cell Culture Supernates, EDTA Plasma, Heparin Plasma, Serum

Intra-Assay Precision Inter-Assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean (pg/mL) 153 744 1419 164 765 1533
Standard Deviation 13.6 26.8 76.6 9.54 61.5 116
CV% 8.9 3.6 5.4 5.8 8.0 7.6

Recovery for Estradiol Parameter Assay Kit

The recovery of Estradiol spiked to levels throughout the range of the assay in various matrices was evaluated.

Sample Type Average % Recovery Range %
Cell Culture Media (n=4) 94 81-102
EDTA Plasma (n=4) 103 90-113
Heparin Plasma (n=4) 102 88-117
Serum (n=4) 96 87-108

Linearity

To assess the linearity of the assay, samples containing and/or spiked with high concentrations of Estradiol were serially diluted with the appropriate Calibrator Diluent to produce samples with values within the dynamic range of the assay. <br>

Multi-species Estradiol ELISA Linearity

Scientific Data Images for Estradiol Parameter Assay Kit

Multi-species Estradiol ELISA Standard Curve

Multi-species Estradiol ELISA Standard Curve

Preparation and Storage

Shipping

The product is shipped at ambient temperature. Upon receipt, store it immediately at the temperature recommended below.

Stability & Storage

Store the unopened product at 2 - 8 °C. Do not use past expiration date.

Background: Estradiol

Alternate Names

Estrogen

Additional Estradiol Products

Product Documents for Estradiol Parameter Assay Kit

Certificate of Analysis

To download a Certificate of Analysis, please enter a lot number in the search box below.

Note: Certificate of Analysis not available for kit components.

Product Specific Notices for Estradiol Parameter Assay Kit

For research use only

Loading...
Loading...
Loading...

⚠ WARNING: This product can expose you to chemicals including N,N-Dimethylforamide, which is known to the State of California to cause cancer. For more information, go to www.P65Warnings.ca.gov.