Human CCL2/MCP-1 ELISA Kit - Quantikine Best Seller
R&D Systems, part of Bio-Techne | Catalog # DCP00
Key Product Details
Assay Length
Sample Type & Volume Required Per Well
Sensitivity
Assay Range
Product Summary for Human CCL2/MCP-1 Quantikine ELISA Kit
Product Specifications
Assay Type
Format
Measurement
Detection Method
Conjugate
Reactivity
Specificity
Cross-reactivity
Interference
Sample Values
Sample Type | Mean (pg/mL) | Range (pg/mL) |
Serum (n=37) | 370 | 200-722 |
EDTA plasma (n=37) | 153 | 72-295 |
Citrate plasma (n=37) | 196 | 134-436 |
Heparin plasma (n=37) | 242 | 113-340 |
Urine (n=37) | 211 | 42-410 |
Cell Culture Supernates - Human peripheral blood leukocytes were cultured in RPMI 1640 and supplemented with 10% fetal bovine serum. The cells were cultured unstimulated or stimulated with 10 μg/mL PHA for 2 or 5 days. Aliquots of the cell culture supernates were removed and assayed for levels of human MCP-1.
Condition | Day 2 (pg/mL) | Day 5 (pg/mL) |
Unstimulated | 647 | 1785 |
Stimulated | 67,225 | 70,000 |
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision.
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision.
Cell Culture Supernates, Urine
Intra-Assay Precision | Inter-Assay Precision | |||||
---|---|---|---|---|---|---|
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 20 | 20 | 20 |
Mean (pg/mL) | 78.2 | 360 | 1079 | 76.0 | 360 | 1086 |
Standard Deviation | 3.3 | 17.7 | 63.5 | 4.5 | 17.2 | 48.5 |
CV% | 4.2 | 4.9 | 5.9 | 5.9 | 4.8 | 4.5 |
Citrate Plasma, EDTA Plasma, Heparin Plasma, Serum
Intra-Assay Precision | Inter-Assay Precision | |||||
---|---|---|---|---|---|---|
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 20 | 20 | 20 |
Mean (pg/mL) | 76.7 | 364 | 1121 | 74.2 | 352 | 1076 |
Standard Deviation | 6.0 | 17.1 | 54.4 | 5.0 | 20.5 | 49.4 |
CV% | 7.8 | 4.7 | 4.9 | 6.7 | 5.8 | 4.6 |
Recovery for Human CCL2/MCP-1 Quantikine ELISA Kit
The recovery of MCP-1 spiked to three levels throughout the range of the assay was evaluated.
Sample Type | Average % Recovery | Range % |
---|---|---|
Cell Culture Media (n=5) | 96 | 88-107 |
Citrate Plasma (n=5) | 100 | 94-107 |
EDTA Plasma (n=5) | 96 | 92-102 |
Heparin Plasma (n=5) | 102 | 94-114 |
Serum (n=5) | 103 | 92-113 |
Urine (n=5) | 92 | 85-100 |
Linearity
To assess the linearity of the assay, samples were spiked with high concentrations of MCP-1 and diluted with the appropriate Calibrator Diluent to produce samples with values within the dynamic range of the assay.
Scientific Data Images for Human CCL2/MCP-1 Quantikine ELISA Kit
Human MCP-1 ELISA/CCL2 ELISA Cell Culture Supernate/Urine Standard Curve
Human MCP-1 ELISA/CCL2 ELISA Serum/Plasma Standard Curve
Preparation and Storage
Shipping
Stability & Storage
Background: CCL2/JE/MCP-1
Alternate Names
Gene Symbol
Additional CCL2/JE/MCP-1 Products
Product Documents for Human CCL2/MCP-1 Quantikine ELISA Kit
Product Specific Notices for Human CCL2/MCP-1 Quantikine ELISA Kit
For research use only
⚠ WARNING: This product can expose you to chemicals including N,N-Dimethylforamide, which is known to the State of California to cause cancer. For more information, go to www.P65Warnings.ca.gov.