Skip to main content

Human CCL2/MCP-1 ELISA Kit - Quantikine Best Seller

R&D Systems, part of Bio-Techne | Catalog # DCP00

R&D Systems, part of Bio-Techne
Catalog #
Availability
Size / Price
Qty
Loading...
DCP00
PDCP00
SCP00

Control Products Available

Key Product Details

Assay Length

3.5 hours or 4.5 hours

Sample Type & Volume Required Per Well

Cell Culture Supernates (200 µL), Serum (100 µL), EDTA Plasma (100 µL), Heparin Plasma (100 µL), Citrate Plasma (100 µL), Urine (100 µL)

Sensitivity

10 pg/mL

Assay Range

31.2-2000 pg/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Citrate Plasma, Urine)

Product Summary for Human CCL2/MCP-1 Quantikine ELISA Kit

The Quantikine Human MCP-1 Immunoassay is a 3.5 or 4.5 hour solid phase ELISA designed to measure MCP-1 in cell culture supernates, serum, plasma, and urine. It contains E. coli-expressed recombinant human MCP-1 and antibodies raised against the recombinant factor. This immunoassay has been shown to accurately quantitate recombinant human MCP-1. Results obtained using natural human MCP-1 showed linear curves that were parallel to the standard curves obtained using the Quantikine kit standards. These results indicate that this kit can be used to determine relative mass values for natural human MCP-1.

Product Specifications

Assay Type

Solid Phase Sandwich ELISA

Format

96-well strip plate

Measurement

Quantitative ELISA

Detection Method

Colorimetric - 450nm (TMB)

Conjugate

HRP

Reactivity

Human

Specificity

Natural and recombinant human MCP-1

Cross-reactivity

< 0.5% cross-reactivity observed with available related molecules. < 50% cross-species reactivity observed with species tested.

Interference

No significant interference observed with available related molecules.

Sample Values

Serum/Plasma/Urine - Samples from apparently healthy volunteers were evaluated for the presence of human MCP-1 in this assay. No medical histories were available for the donors used in this study. The reported urine values are actual and are not normalized for creatinine content.

Sample TypeMean (pg/mL)Range (pg/mL)
Serum (n=37)370200-722
EDTA plasma (n=37)15372-295
Citrate plasma (n=37)196134-436
Heparin plasma (n=37)242113-340
Urine (n=37)21142-410

Cell Culture Supernates - Human peripheral blood leukocytes were cultured in RPMI 1640 and supplemented with 10% fetal bovine serum. The cells were cultured unstimulated or stimulated with 10 μg/mL PHA for 2 or 5 days. Aliquots of the cell culture supernates were removed and assayed for levels of human MCP-1.

ConditionDay 2 (pg/mL)Day 5 (pg/mL)
Unstimulated6471785
Stimulated67,22570,000


Precision

Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision.

Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision.

Cell Culture Supernates, Urine

Intra-Assay Precision Inter-Assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean (pg/mL) 78.2 360 1079 76.0 360 1086
Standard Deviation 3.3 17.7 63.5 4.5 17.2 48.5
CV% 4.2 4.9 5.9 5.9 4.8 4.5

Citrate Plasma, EDTA Plasma, Heparin Plasma, Serum

Intra-Assay Precision Inter-Assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean (pg/mL) 76.7 364 1121 74.2 352 1076
Standard Deviation 6.0 17.1 54.4 5.0 20.5 49.4
CV% 7.8 4.7 4.9 6.7 5.8 4.6

Recovery for Human CCL2/MCP-1 Quantikine ELISA Kit

The recovery of MCP-1 spiked to three levels throughout the range of the assay was evaluated.

Sample Type Average % Recovery Range %
Cell Culture Media (n=5) 96 88-107
Citrate Plasma (n=5) 100 94-107
EDTA Plasma (n=5) 96 92-102
Heparin Plasma (n=5) 102 94-114
Serum (n=5) 103 92-113
Urine (n=5) 92 85-100

Linearity

To assess the linearity of the assay, samples were spiked with high concentrations of MCP-1 and diluted with the appropriate Calibrator Diluent to produce samples with values within the dynamic range of the assay.

Human MCP-1 ELISA/CCL2 ELISA Linearity

Scientific Data Images for Human CCL2/MCP-1 Quantikine ELISA Kit

Human MCP-1 ELISA/CCL2 ELISA Cell Culture Supernate/Urine Standard Curve

Human MCP-1 ELISA/CCL2 ELISA Cell Culture Supernate/Urine Standard Curve

Human MCP-1 ELISA/CCL2 ELISA Serum/Plasma Standard Curve

Human MCP-1 ELISA/CCL2 ELISA Serum/Plasma Standard Curve

Preparation and Storage

Shipping

The product is shipped at ambient temperature. Upon receipt, store it immediately at the temperature recommended below.

Stability & Storage

Store the unopened product at 2 - 8 °C. Do not use past expiration date.

Background: CCL2/JE/MCP-1

CCL2/JE/MCP-1, is a chemokine that binds the receptor CCR2 and induces the chemoattraction of mononuclear cells. It induces the activation of monocytes, NK cells, lymphocytes, and basophils. Additionally, CCL2 promotes Th2 polarization in CD4+ T cells, and CCL2-mediated recruitment of monocytes to sites of inflammation contributes to disease severity in atherosclerosis, multiple sclerosis, and allergic asthma. Endogenous proteolytic trimming of CCL2 at the N-terminus, including the N-terminal pyrrolidone carboxylic acid-modified glutamine, downregulates activity but not receptor binding.

Alternate Names

GDCF-2, HC11, HSMCR30, MCAF, MCP-1, SMC-CF

Entrez Gene IDs

6347 (Human); 20296 (Mouse); 24770 (Rat); 403981 (Canine)

Gene Symbol

CCL2

Additional CCL2/JE/MCP-1 Products

Product Documents for Human CCL2/MCP-1 Quantikine ELISA Kit

Certificate of Analysis

To download a Certificate of Analysis, please enter a lot number in the search box below.

Note: Certificate of Analysis not available for kit components.

Product Specific Notices for Human CCL2/MCP-1 Quantikine ELISA Kit

For research use only

Loading...
Loading...
Loading...

⚠ WARNING: This product can expose you to chemicals including N,N-Dimethylforamide, which is known to the State of California to cause cancer. For more information, go to www.P65Warnings.ca.gov.