Human CXCL13/BLC/BCA-1 Quantikine ELISA Kit Best Seller
R&D Systems, part of Bio-Techne | Catalog # DCX130
Key Product Details
Assay Length
Sample Type & Volume Required Per Well
Sensitivity
Assay Range
Product Summary for Human CXCL13/BLC/BCA-1 Quantikine ELISA Kit
Product Specifications
Assay Type
Format
Measurement
Detection Method
Conjugate
Reactivity
Specificity
Cross-reactivity
Interference
Sample Values
Sample Type | Mean (pg/mL) | Range (pg/mL) | Standard Deviation (pg/mL) |
Serum* (n=35) | 81.9 | 39.4-252 | 43.5 |
EDTA plasma (n=35) | 62.5 | 29.4-246 | 42.7 |
Heparin plasma (n=35) | 63.5 | 21.7-240 | 44.1 |
Sample Type | Mean of Detectable (pg/mL) | % Detectable | Range (pg/mL) |
Saliva (n=10) | 67.4 | 70 | ND-284 |
Condition | Day 1 (pg/mL) | Day 5 (pg/mL) |
Unstimulated | ND | ND |
Stimulated | 11 | 640 |
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision.
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision.
Cell Culture Supernates, EDTA Plasma, Heparin Plasma, Saliva, Serum
Intra-Assay Precision | Inter-Assay Precision | |||||
---|---|---|---|---|---|---|
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 40 | 40 | 40 |
Mean (pg/mL) | 50.2 | 136 | 294 | 54.0 | 138 | 285 |
Standard Deviation | 2.2 | 4.5 | 8.0 | 5.1 | 13.2 | 24.8 |
CV% | 4.3 | 3.3 | 2.7 | 9.4 | 9.6 | 8.7 |
Recovery for Human CXCL13/BLC/BCA-1 Quantikine ELISA Kit
The recovery of natural BLC/BCA-1 spiked to levels throughout the range of the assay in various matrices was evaluated.
Sample Type | Average % Recovery | Range % |
---|---|---|
Cell Culture Media (n=4) | 98 | 87-110 |
EDTA Plasma (n=4) | 99 | 92-107 |
Heparin Plasma (n=4) | 101 | 92-115 |
Serum (n=4) | 99 | 86-108 |
Linearity
To assess the linearity of the assay, samples containing and/or spiked with high concentrations of natural BLC/BCA-1 were serially diluted with the Calibrator Diluent to produce samples with values within the dynamic range of the assay.
Scientific Data Images for Human CXCL13/BLC/BCA-1 Quantikine ELISA Kit
Human CXCL13/BLC/BCA-1 ELISA Standard Curve
Human CXCL13/BLC/BCA-1 Quantikine ELISA Kit
Correlation between European League Against Rheumatism Sjögren’s Syndrome Disease Activity Index (ESSDAI) scores and concentrations of nine screened serum proteins analyzed by ELISA in the validation cohort of patients with primary Sjögren’s syndrome (n = 58). The nine proteins were CXCL13, TNF receptor 2 (TNF-R2), CD48, B-cell activating factor (BAFF), programmed cell death protein 1 ligand 2 (PD-L2), lymphocyte activation gene-3 (LAG-3), beta2-microglobulin ( beta2MG), ephrin type-B receptor 2 (EPHB2), and CD163. Pearson’s correlation test was used. The correlation coefficients (r) and P values are shown in the table. P <0.05 was considered significant Image collected and cropped by CiteAb from the following open publication (https://arthritis-research.biomedcentral.com/articles/10.1186/s13075-016-1006-1), licensed under a CC-BY license. Not internally tested by R&D Systems.Preparation and Storage
Shipping
Stability & Storage
Background: CXCL13/BLC/BCA-1
Alternate Names
Gene Symbol
Additional CXCL13/BLC/BCA-1 Products
Product Documents for Human CXCL13/BLC/BCA-1 Quantikine ELISA Kit
Product Specific Notices for Human CXCL13/BLC/BCA-1 Quantikine ELISA Kit
For research use only
⚠ WARNING: This product can expose you to chemicals including N,N-Dimethylforamide, which is known to the State of California to cause cancer. For more information, go to www.P65Warnings.ca.gov.