Human HE4/WFDC2 Quantikine ELISA Kit
R&D Systems, part of Bio-Techne | Catalog # DHE400
Key Product Details
Assay Length
Sample Type & Volume Required Per Well
Sensitivity
Assay Range
Product Summary for Human HE4/WFDC2 Quantikine ELISA Kit
Product Specifications
Assay Type
Format
Measurement
Detection Method
Conjugate
Reactivity
Specificity
Cross-reactivity
Interference
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision.
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision.
Cell Culture Supernates, EDTA Plasma, Heparin Plasma, Saliva, Serum
Intra-Assay Precision | Inter-Assay Precision | |||||
---|---|---|---|---|---|---|
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 20 | 20 | 20 |
Mean (pg/mL) | 470 | 1278 | 2448 | 476 | 1408 | 2731 |
Standard Deviation | 11.5 | 35.3 | 86.1 | 26.0 | 79.1 | 162 |
CV% | 2.4 | 2.8 | 3.5 | 5.5 | 5.6 | 5.9 |
Urine
Intra-Assay Precision | Inter-Assay Precision | |||||
---|---|---|---|---|---|---|
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 20 | 20 | 20 |
Mean (pg/mL) | 451 | 1337 | 2629 | 452 | 1409 | 2761 |
Standard Deviation | 14.0 | 39.8 | 85.7 | 25.0 | 78.0 | 202 |
CV% | 3.1 | 3.0 | 3.3 | 5.5 | 5.5 | 7.3 |
Recovery for Human HE4/WFDC2 Quantikine ELISA Kit
The recovery of HE4/WFDC2 spiked to levels throughout the range of the assay in various matrices was evaluated.
Sample Type | Average % Recovery | Range % |
---|---|---|
Cell Culture Media (n=8) | 103 | 95-114 |
EDTA Plasma (n=4) | 108 | 94-115 |
Heparin Plasma (n=4) | 107 | 96-115 |
Saliva (n=4) | 100 | 98-107 |
Serum (n=4) | 104 | 92-115 |
Urine (n=4) | 98 | 85-111 |
Linearity
To assess the linearity of the assay, samples containing high concentrations of HE4/WFDC2 were serially diluted with calibrator diluent to produce samples with values within the dynamic range of the assay.
Scientific Data Images for Human HE4/WFDC2 Quantikine ELISA Kit
Human HE4/WFDC2 ELISA Cell Culture Supernate/Serum/Plasma/Saliva Standard Curve
Human HE4/WFDC2 ELISA Urine Standard Curve
Preparation and Storage
Shipping
Stability & Storage
Background: HE4/WFDC2
Long Name
Alternate Names
Gene Symbol
Additional HE4/WFDC2 Products
Product Documents for Human HE4/WFDC2 Quantikine ELISA Kit
Product Specific Notices for Human HE4/WFDC2 Quantikine ELISA Kit
For research use only
⚠ WARNING: This product can expose you to chemicals including N,N-Dimethylforamide, which is known to the State of California to cause cancer. For more information, go to www.P65Warnings.ca.gov.