Human IFN-beta Quantikine ELISA Kit Best Seller
R&D Systems, part of Bio-Techne | Catalog # DIFNB0
Key Product Details
Assay Length
Sample Type & Volume Required Per Well
Sensitivity
Assay Range
Product Summary for Human IFN-beta Quantikine ELISA Kit
Product Specifications
Measurement
Detection Method
Conjugate
Reactivity
Specificity
Cross-reactivity
Interference
Sample Values
Condition | pg/mL |
Unstimulated | ND |
Stimulated | 848 |
Condition | pg/mL |
Unstimulated | ND |
Stimulated | 2323 |
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in twenty separate assays to assess inter-assay precision. Assays were performed by at least three technicians using two lots of components.
Cell Culture Supernates
Intra-Assay Precision | Inter-Assay Precision | |||||
---|---|---|---|---|---|---|
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 20 | 20 | 20 |
Mean (pg/mL) | 58.3 | 159 | 295 | 56.4 | 160 | 275 |
Standard Deviation | 1.71 | 6.01 | 13.8 | 5.26 | 13.0 | 21.9 |
CV% | 2.9 | 3.8 | 4.7 | 9.3 | 8.1 | 8.0 |
Recovery for Human IFN-beta Quantikine ELISA Kit
The recovery of human IFN-beta spiked to levels throughout the range of the assay in various matrices was evaluated.
Sample Type | Average % Recovery | Range % |
---|---|---|
Cell Culture Supernates (n=5) | 107 | 95-111 |
Linearity
To assess the linearity of the assay, samples containing high concentrations of human IFN-beta were serially diluted with calibrator diluent to produce samples with values within the dynamic range of the assay.
Scientific Data Images for Human IFN-beta Quantikine ELISA Kit
Human IFN-beta Quantikine ELISA Spiked Recovery Competitor Comparision
IFN-beta is spiked at three known concentrations throughout the range of the assay and run to measure response of the spiked sample matrix. R&D Systems recovery is 98% compared to 55% for the top competitor.Human IFN-beta Quantikine ELISA Natural Linearity Competitor Comparison
Samples containing high concentrations of IFN-beta in various matrices are diluted with appropriate Calibrator Diluent to produce samples with values within the dynamic range of the assay. R&D Systems linearity is between 103%-112% compared to 119%-170% for the top competitor.Human IFN- beta ELISA Standard Curve
Preparation and Storage
Shipping
Stability & Storage
Background: IFN-beta
Long Name
Alternate Names
Gene Symbol
Additional IFN-beta Products
Product Documents for Human IFN-beta Quantikine ELISA Kit
Product Specific Notices for Human IFN-beta Quantikine ELISA Kit
For research use only
⚠ WARNING: This product can expose you to chemicals including N,N-Dimethylforamide, which is known to the State of California to cause cancer. For more information, go to www.P65Warnings.ca.gov.