Human SPARC Quantikine ELISA Kit
R&D Systems, part of Bio-Techne | Catalog # DSP00
Key Product Details
Assay Length
Sample Type & Volume Required Per Well
Sensitivity
Assay Range
Product Summary for Human SPARC Quantikine ELISA Kit
Product Specifications
Assay Type
Format
Measurement
Detection Method
Conjugate
Reactivity
Specificity
Cross-reactivity
Interference
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision.
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision.
Cell Culture Supernates, Human Milk, Platelet-poor EDTA Plasma, Platelet-poor Heparin Plasma, Serum
Intra-Assay Precision | Inter-Assay Precision | |||||
---|---|---|---|---|---|---|
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 20 | 20 | 20 |
Mean (ng/mL) | 7.34 | 14.3 | 29.9 | 6.65 | 13.5 | 28.3 |
Standard Deviation | 0.129 | 0.282 | 0.733 | 0.541 | 0.989 | 2.40 |
CV% | 1.8 | 2.0 | 2.5 | 8.1 | 7.3 | 8.5 |
Recovery for Human SPARC Quantikine ELISA Kit
The recovery of SPARC spiked to levels throughout the range of the assay in various matrices was evaluated.
Sample Type | Average % Recovery | Range % |
---|---|---|
Cell Culture Media (n=4) | 102 | 86-111 |
Platelet-poor EDTA Plasma (n=4) | 104 | 93-111 |
Platelet-poor Heparin Plasma (n=4) | 91 | 82-100 |
Linearity
To assess the linearity of the assay, samples containing and/or spiked with high concentrations of SPARC were serially diluted with Calibrator Diluent to produce samples with values within the dynamic range of the assay.
Scientific Data Images for Human SPARC Quantikine ELISA Kit
Human SPARC ELISA Standard Curve
Preparation and Storage
Shipping
Stability & Storage
Background: SPARC
Long Name
Alternate Names
Gene Symbol
Additional SPARC Products
Product Documents for Human SPARC Quantikine ELISA Kit
Product Specific Notices for Human SPARC Quantikine ELISA Kit
For research use only
⚠ WARNING: This product can expose you to chemicals including N,N-Dimethylforamide, which is known to the State of California to cause cancer. For more information, go to www.P65Warnings.ca.gov.