Mouse CCL2/JE/MCP-1 Quantikine ELISA Kit Best Seller

R&D Systems, part of Bio-Techne | Catalog # MJE00B

R&D Systems, part of Bio-Techne
Catalog #
Availability
Size / Price
Qty
Loading...
MJE00B
PMJE00B
SMJE00B

Key Product Details

Assay Length

4.5 hours

Sample Type & Volume Required Per Well

Cell Culture Supernates (50 µL), Tissue Lysates (50 µL), Tissue Culture Supernates (50 µL), Serum (50 µL), EDTA Plasma (50 µL), Heparin Plasma (50 µL)

Sensitivity

0.666 pg/mL

Assay Range

7.81-500 pg/mL (Cell Culture Supernates, Tissue Lysates, Tissue Culture Supernates, Serum, EDTA Plasma, Heparin Plasma)

Product Summary for Mouse CCL2/JE/MCP-1 Quantikine ELISA Kit

The Quantikine® Mouse CCL2/JE/MCP-1 Immunoassay is a 4.5 hour solid-phase ELISA designed to measure mouse MCP-1 in cell culture supernates, tissue culture supernates, tissue lysates, serum, and plasma. It contains E. coli-expressed recombinant mouse MCP-1 and antibodies raised against the recombinant factor. This immunoassay has been shown to accurately quantitate the recombinant mouse MCP-1. Results obtained using natural mouse MCP-1 showed dose response curves that were parallel to the standard curves obtained using the Quantikine® kit standards. These results indicate that this kit can be used to determine relative mass values for natural mouse MCP-1.

Product Specifications

Measurement

Quantitative ELISA

Detection Method

Colorimetric - 450nm (TMB)

Conjugate

HRP

Reactivity

Mouse

Specificity

Natural and recombinant mouse MCP-1.

Cross-reactivity

< 0.5% cross-reactivity observed with available related molecules. < 50% cross-species reactivity observed with species tested.

Interference

No significant interference observed with available related molecules.

Precision

Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in twenty separate assays to assess inter-assay precision. Assays were performed by at least three technicians.

Cell Culture Supernates, EDTA Plasma, Heparin Plasma, Serum, Tissue Culture Supernates, Tissue Lysates

Intra-Assay Precision Inter-Assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean (pg/mL) 36.7 51.8 149 37.8 60.0 155
Standard Deviation 1.45 1.40 3.66 2.85 4.37 7.86
CV% 4.0 2.7 2.5 7.5 7.3 5.1

Recovery for Mouse CCL2/JE/MCP-1 Quantikine ELISA Kit

The recovery of mouse MCP-1 spiked to three levels throughout the range of the assay in various matrices was evaluated.

Sample Type Average % Recovery Range %
Cell Culture Media (n=4) 105 95-112
Cell Lysis Buffer (n=4) 88 79-101
EDTA Plasma (n=4) 84 75-93
Heparin Plasma (n=4) 93 80-101
Serum (n=4) 91 83-100

Linearity

To assess the linearity of the assay, samples containing and/or spiked with high concentrations of mouse MCP-1 were serially diluted with calibrator diluent to produce samples with values within the dynamic range of the assay.

Mouse/Rat CCL2/JE/MCP-1 ELISA Linearity
Mouse CCL2/JE/MCP-1 ELISA Linearity

Scientific Data Images for Mouse CCL2/JE/MCP-1 Quantikine ELISA Kit

Mouse CCL2/JE/MCP-1 ELISA Standard Curve

Mouse CCL2/JE/MCP-1 ELISA Standard Curve

Mouse/Rat CCL2/JE/MCP-1 Standard Curve

Mouse/Rat CCL2/JE/MCP-1 Standard Curve

Preparation and Storage

Shipping

The product is shipped at ambient temperature. Upon receipt, store it immediately at the temperature recommended below.

Stability & Storage

Store the unopened product at 2 - 8 °C. Do not use past expiration date.

Background: CCL2/JE/MCP-1

CCL2/JE/MCP-1, is a chemokine that binds the receptor CCR2 and induces the chemoattraction of mononuclear cells. It induces the activation of monocytes, NK cells, lymphocytes, and basophils. Additionally, CCL2 promotes Th2 polarization in CD4+ T cells, and CCL2-mediated recruitment of monocytes to sites of inflammation contributes to disease severity in atherosclerosis, multiple sclerosis, and allergic asthma. Endogenous proteolytic trimming of CCL2 at the N-terminus, including the N-terminal pyrrolidone carboxylic acid-modified glutamine, downregulates activity but not receptor binding.

Alternate Names

GDCF-2, HC11, HSMCR30, MCAF, MCP-1, SMC-CF

Entrez Gene IDs

6347 (Human); 20296 (Mouse); 24770 (Rat); 403981 (Canine)

Gene Symbol

CCL2

Additional CCL2/JE/MCP-1 Products

Product Documents for Mouse CCL2/JE/MCP-1 Quantikine ELISA Kit

Certificate of Analysis

To download a Certificate of Analysis, please enter a lot number in the search box below.

Note: Certificate of Analysis not available for kit components.

Product Specific Notices for Mouse CCL2/JE/MCP-1 Quantikine ELISA Kit

For research use only

Loading...
Loading...
Loading...

⚠ WARNING: This product can expose you to chemicals including N,N-Dimethylforamide, which is known to the State of California to cause cancer. For more information, go to www.P65Warnings.ca.gov.