Skip to main content

Human Coagulation Factor III/Tissue Factor Quantikine ELISA

R&D Systems, part of Bio-Techne | Catalog # DCF300

R&D Systems, part of Bio-Techne
Catalog #
Availability
Size / Price
Qty
Loading...
DCF300

Control Products Available

Key Product Details

Assay Length

4.5 hours

Sample Type & Volume Required Per Well

Cell Culture Supernates (50 µL), Cell Lysates (50 µL), EDTA Plasma (50 µL), Heparin Plasma (50 µL), Citrate Plasma (50 µL), Urine (50 µL)

Sensitivity

2.05 pg/mL

Assay Range

7.8-500 pg/mL (Cell Culture Supernates, Cell Lysates, EDTA Plasma, Heparin Plasma, Citrate Plasma, Urine)

Product Summary for Human Coagulation Factor III/Tissue Factor Quantikine ELISA

The Quantikine Human Coagulation Factor III/Tissue Factor Immunoassay is a 4.5 hour solid-phase ELISA designed to measure human Coagulation Factor III in cell culture supernates, cell lysates, plasma, and urine. It contains NS0-expressed recombinant human Coagulation Factor III and has been shown to accurately quantitate the recombinant factor. Results obtained using natural human Coagulation Factor III showed linear curves that were p arallel to the standard curves obtained using the Quantikine kit standards. These results indicate that this kit can be used to determine relative mass values for naturally occurring Coagulation Factor III.

Product Specifications

Assay Type

Solid Phase Sandwich ELISA

Format

96-well strip plate

Measurement

Quantitative ELISA

Detection Method

Colorimetric - 450nm (TMB)

Conjugate

HRP

Reactivity

Human

Specificity

Natural and recombinant human Coagulation Factor III

Cross-reactivity

< 0.5% cross-reactivity observed with available related molecules. < 50% cross-species reactivity observed with species tested.

Interference

No significant interference observed with available related molecules.

Precision

Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision.

Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision.

Cell Culture Supernates, Cell Lysates, Citrate Plasma, EDTA Plasma, Heparin Plasma, Urine

Intra-Assay Precision Inter-Assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 40 40 40
Mean (pg/mL) 33.4 65.4 148 71 144 301
Standard Deviation 0.78 1.81 5.0 4.5 7.8 17.1
CV% 2.3 2.8 3.4 6.3 5.4 5.7

Recovery for Human Coagulation Factor III/Tissue Factor Quantikine ELISA

The recovery of Coagulation Factor III spiked to levels throughout the range of the assay in various matrices was evaluated.

Sample Type Average % Recovery Range %
Cell Culture Media (n=4) 104 95-110
Citrate Plasma (n=4) 92 86-102
EDTA Plasma (n=4) 91 86-97
Heparin Plasma (n=4) 95 88-109
Urine (n=4) 102 96-112

Linearity

To assess the linearity of the assay, samples containing and/or spiked with high concentrations of Coagulation Factor III were serially diluted with the Calibrator Diluent to produce samples with values within the dynamic range of the assay.

Human Coagulation Factor III/Tissue Factor ELISA Linearity

Scientific Data Images for Human Coagulation Factor III/Tissue Factor Quantikine ELISA

Human Coagulation Factor III/Tissue Factor ELISA Standard Curve

Human Coagulation Factor III/Tissue Factor ELISA Standard Curve

Preparation and Storage

Shipping

The product is shipped at ambient temperature. Upon receipt, store it immediately at the temperature recommended below.

Stability & Storage

Store the unopened product at 2 - 8 °C. Do not use past expiration date.

Background: Coagulation Factor III/Tissue Factor

Coagulation Factor III, also known as Tissue Factor (TF), Thromboplastin, and CD142, is the primary initiator of the extrinsic coagulation pathway. It is a transmembrane protein that is consitutively expressed in subendothelial cells throughout the vasculature and is inducible on endothelial cells and monocytes. Damage to the vessel wall exposes Coagulation Factor III to circulating Factor VII, triggering the proteolytic cascade that results in fibrin clot formation. Coagulation Factor III plays an important role in thrombotic diseases, disseminated intravascular coagulation, and atherosclerosis. It also plays a role in non-hemostatic inflammation, angiogenesis, and tumor progression.

Alternate Names

CD142, F3, Thromboplastin, Tissue Factor

Entrez Gene IDs

2152 (Human); 14066 (Mouse)

Gene Symbol

F3

Additional Coagulation Factor III/Tissue Factor Products

Product Documents for Human Coagulation Factor III/Tissue Factor Quantikine ELISA

Certificate of Analysis

To download a Certificate of Analysis, please enter a lot number in the search box below.

Note: Certificate of Analysis not available for kit components.

Product Specific Notices for Human Coagulation Factor III/Tissue Factor Quantikine ELISA

For research use only

Loading...
Loading...
Loading...

⚠ WARNING: This product can expose you to chemicals including N,N-Dimethylforamide, which is known to the State of California to cause cancer. For more information, go to www.P65Warnings.ca.gov.