Human CXCL4/PF4 Quantikine ELISA Kit
R&D Systems, part of Bio-Techne | Catalog # DPF40
Key Product Details
Assay Length
Sample Type & Volume Required Per Well
Sensitivity
Assay Range
Product Summary for Human CXCL4/PF4 Quantikine ELISA Kit
Product Specifications
Assay Type
Format
Measurement
Detection Method
Conjugate
Reactivity
Specificity
Cross-reactivity
Interference
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in twenty separate assays to assess inter-assay precision. Assays were performed by at least three technicians using two lots of components.
Cell Culture Supernates, Platelet-poor EDTA Plasma, Platelet-poor Heparin Plasma, Serum
Intra-Assay Precision | Inter-Assay Precision | |||||
---|---|---|---|---|---|---|
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 20 | 20 | 20 |
Mean (ng/mL) | 5.80 | 14.1 | 26.8 | 5.61 | 14.3 | 27.2 |
Standard Deviation | 0.472 | 0.840 | 1.85 | 0.558 | 1.70 | 3.20 |
CV% | 8.1 | 6.0 | 6.9 | 9.9 | 11.9 | 11.8 |
Recovery for Human CXCL4/PF4 Quantikine ELISA Kit
The recovery of human PF4 spiked to levels throughout the range of the assay in various matrices was evaluated.
Sample Type | Average % Recovery | Range % |
---|---|---|
Cell Culture Media (n=4) | 110 | 99-117 |
Platelet-poor EDTA Plasma (n=4) | 100 | 89-117 |
Linearity
To assess linearity of the assay, samples containing and/or spiked with high concentrations of human PF4 were diluted with Calibrator Diluent to produce samples with values within the dynamic range of the assay.
Scientific Data Images for Human CXCL4/PF4 Quantikine ELISA Kit
Human CXCL4/PF4 ELISA Standard Curve
Preparation and Storage
Shipping
Stability & Storage
Background: CXCL4/PF4
Additional CXCL4/PF4 Products
Product Documents for Human CXCL4/PF4 Quantikine ELISA Kit
Product Specific Notices for Human CXCL4/PF4 Quantikine ELISA Kit
For research use only
⚠ WARNING: This product can expose you to chemicals including N,N-Dimethylforamide, which is known to the State of California to cause cancer. For more information, go to www.P65Warnings.ca.gov.