Human Leptin ELISA Kit - Quantikine Best Seller
R&D Systems, part of Bio-Techne | Catalog # DLP00
Key Product Details
Assay Length
Sample Type & Volume Required Per Well
Sensitivity
Assay Range
Product Summary for Human Leptin Quantikine ELISA Kit
Product Specifications
Assay Type
Format
Measurement
Detection Method
Conjugate
Reactivity
Specificity
Cross-reactivity
Interference
Sample Values
Sample Type | Range (pg/mL) | Mean (pg/mL) |
Male Serum (n=16) | 2205-11,419 | 4760 |
Female Serum (n=36) | 3877-77,273 | 20,676 |
Five additional male serum samples fell below the lowest standard,15.6 pg/mL, when diluted 100-fold.
Condition | Day 1(pg/mL) | Day 5 (pg/mL) |
Unstimulated | ND | ND |
Stimulated | 152 | 75.9 |
Condition | Day 1 (pg/mL) | Day 2 (pg/mL) | Day 3 (pg/mL) |
Unstimulated* | 849 | 1549 | 1667 |
2 μM forskolin* | 1231 | 1699 | 2054 |
20 μM forskolin* | 1137 | 1725 | 2747 |
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision.
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision.
Cell Culture Supernates, Citrate Plasma, EDTA Plasma, Heparin Plasma, Serum
Intra-Assay Precision | Inter-Assay Precision | |||||
---|---|---|---|---|---|---|
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 40 | 40 | 40 |
Mean (pg/mL) | 64.5 | 146 | 621 | 65.7 | 146 | 581 |
Standard Deviation | 2.14 | 4.32 | 20.0 | 3.56 | 6.17 | 20.6 |
CV% | 3.3 | 3.0 | 3.2 | 5.4 | 4.2 | 3.5 |
Recovery for Human Leptin Quantikine ELISA Kit
The recovery of Leptin spiked to three different levels throughout the range of the assay in various matrices was evaluated.
Sample Type | Average % Recovery | Range % |
---|---|---|
Cell Culture Media (n=5) | 98 | 94-102 |
Citrate Plasma (n=5) | 95 | 87-105 |
EDTA Plasma (n=5) | 99 | 85-112 |
Heparin Plasma (n=5) | 90 | 81-100 |
Serum (n=5) | 95 | 89-109 |
Linearity
To assess the linearity of the assay, cell culture media samples spiked with high concentrations of Leptin and natural serum/plasma samples were diluted with Calibrator Diluent to produce samples with values within the dynamic range of the assay.
Scientific Data Images for Human Leptin Quantikine ELISA Kit
Human Leptin/OB ELISA Standard Curve
Preparation and Storage
Shipping
Stability & Storage
Background: Leptin/OB
Additional Leptin/OB Products
Product Documents for Human Leptin Quantikine ELISA Kit
Product Specific Notices for Human Leptin Quantikine ELISA Kit
For research use only
⚠ WARNING: This product can expose you to chemicals including N,N-Dimethylforamide, which is known to the State of California to cause cancer. For more information, go to www.P65Warnings.ca.gov.