Human TNF RI/TNFRSF1A Quantikine QuicKit ELISA
R&D Systems, part of Bio-Techne | Catalog # QK225
Key Product Details
Assay Length
Sample Type & Volume Required Per Well
Sensitivity
Assay Range
Product Summary for Human TNF RI/TNFRSF1A Quantikine QuicKit ELISA
Product Specifications
Measurement
Detection Method
Conjugate
Reactivity
Specificity
Cross-reactivity
Interference
Sample Values
Sample Type | Mean (pg/mL) | Range (pg/mL) | Standard (pg/mL) |
Serum (n=10) | 1341 | 1050-1948 | 257 |
EDTA Plasma (n=10) | 1092 | 849-1636 | 233 |
Heparin Plasma (n=10) | 1102 | 861-1600 | 251 |
Urine (n=10) | 1408 | 836-2140 | 485 |
Condition | Day 5 (pg/mL) |
Unstimulated | 172 |
Stimulated with PHA | 280 |
Stimulated with Concanavalin A | 234 |
Precision
Intra-Assay Precision (Precision within an assay) Two samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-Assay Precision (Precision between assays) Two samples of known concentration were tested in ten separate assays to assess inter-assay precision. Assays were performed by at least three technicians.
Cell Culture Supernates, EDTA Plasma, Heparin Plasma, Serum, Urine
Intra-Assay Precision | Inter-Assay Precision | |||
---|---|---|---|---|
Sample | 1 | 2 | 1 | 1 |
n | 20 | 20 | 10 | 10 |
Mean (pg/mL) | 100 | 624 | 99.9 | 619 |
Standard Deviation | 1.75 | 17.0 | 6.78 | 39.3 |
CV% | 1.8 | 2.7 | 6.8 | 6.3 |
Recovery for Human TNF RI/TNFRSF1A Quantikine QuicKit ELISA
The recovery of human TNF RI spiked to three different levels in samples throughout the range of the assay in various matrices was evaluated.
Sample Type | Average % Recovery | Range % |
---|---|---|
Cell Culture Media (n=4) | 101 | 92-109 |
EDTA Plasma (n=2) | 98 | 88-109 |
Heparin Plasma (n=2) | 98 | 85-109 |
Serum (n=2) | 99 | 91-113 |
Urine (n=2) | 96 | 92-100 |
Linearity
To assess the linearity of the assay, samples containing and/or spiked with high concentrations of human TNF RI were diluted with calibrator diluent to produce samples with values within the dynamic range of the assay.
Scientific Data Images for Human TNF RI/TNFRSF1A Quantikine QuicKit ELISA
Human TNF RI ELISA Standard Curve
Human TNF RI QuicKit Spiked Recovery Competitor Comparison
TNF RI is spiked at three known concentrations throughout the range of the assay and run to measure response of the spiked sample matrix. Serum recovery is 108% compared to 81% for the top competitor. Plasma recovery is 82% compared to 75% for the top competitor. Urine recovery is 92% compared to 83% for the top competitor. In spike and recovery experiments, natural samples are spiked with the recombinant target analyte of interest to identify interference caused by sample matrices.Human TNF RI QuicKit Natural Linearity Competitor Comparison
Samples containing TNF RI were diluted with appropriate Calibrator Diluent to produce samples with values within the dynamic range of the assay. The linearity is between 108%-117% compared to 113%-164% for the top competitor.Preparation and Storage
Shipping
Stability & Storage
Background: TNF RI/TNFRSF1A
Long Name
Alternate Names
Gene Symbol
Additional TNF RI/TNFRSF1A Products
Product Documents for Human TNF RI/TNFRSF1A Quantikine QuicKit ELISA
Product Specific Notices for Human TNF RI/TNFRSF1A Quantikine QuicKit ELISA
For research use only