Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision.
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision.
Cell Culture Supernates, EDTA Plasma, Heparin Plasma, Serum
Intra-Assay Precision | Inter-Assay Precision | |||||
---|---|---|---|---|---|---|
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 24 | 24 | 24 |
Mean (pg/mL) | 19.9 | 51.8 | 382 | 20.0 | 52.6 | 383 |
Standard Deviation | 1.3 | 2.5 | 19.2 | 1.7 | 2.9 | 28.7 |
CV% | 6.5 | 4.8 | 5.0 | 8.5 | 5.5 | 7.5 |
Recovery for Mouse CCL22/MDC Quantikine ELISA Kit
The recovery of mouse MDC spiked to three levels throughout the range of the assay in various matrices was evaluated.
Sample Type | Average % Recovery | Range % |
---|---|---|
Cell Culture Supernates (n=7) | 102 | 87-112 |
EDTA Plasma (n=4) | 94 | 89-98 |
Heparin Plasma (n=4) | 90 | 80-109 |
Serum (n=5) | 91 | 82-100 |
Linearity
To assess the linearity of the assay, samples containing and/or spiked with high concentrations of mouse MDC in each matrix were diluted with Calibrator Diluent and then assayed.
Scientific Data Images for Mouse CCL22/MDC Quantikine ELISA Kit
Mouse CCL22/MDC ELISA Standard Curve
Preparation and Storage
Shipping
Stability & Storage
Background: CCL22/MDC
Additional CCL22/MDC Products
Product Documents for Mouse CCL22/MDC Quantikine ELISA Kit
Product Specific Notices for Mouse CCL22/MDC Quantikine ELISA Kit
For research use only
⚠ WARNING: This product can expose you to chemicals including N,N-Dimethylforamide, which is known to the State of California to cause cancer. For more information, go to www.P65Warnings.ca.gov.