Mouse CXCL13/BLC/BCA-1 Quantikine ELISA Kit
R&D Systems, part of Bio-Techne | Catalog # MCX130
Key Product Details
Assay Length
Sample Type & Volume Required Per Well
Sensitivity
Assay Range
Product Summary for Mouse CXCL13/BLC/BCA-1 Quantikine ELISA Kit
Product Specifications
Assay Type
Format
Measurement
Detection Method
Conjugate
Reactivity
Specificity
Cross-reactivity
Interference
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision.
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision.
Cell Culture Supernates, EDTA Plasma, Heparin Plasma, Serum
Intra-Assay Precision | Inter-Assay Precision | |||||
---|---|---|---|---|---|---|
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 20 | 20 | 20 |
Mean (pg/mL) | 51 | 109 | 408 | 52 | 104 | 366 |
Standard Deviation | 3.1 | 3.8 | 15.4 | 5.1 | 7.1 | 27.1 |
CV% | 6.1 | 3.5 | 3.8 | 9.8 | 6.8 | 7.4 |
Recovery for Mouse CXCL13/BLC/BCA-1 Quantikine ELISA Kit
The recovery of mouse BLC/BCA-1 spiked to three levels throughout the range of the assay in various matrices was evaluated.
Sample Type | Average % Recovery | Range % |
---|---|---|
Cell Culture Supernates (n=4) | 93 | 83-109 |
EDTA Plasma (n=7) | 98 | 80-115 |
Heparin Plasma (n=4) | 96 | 80-110 |
Serum (n=7) | 104 | 85-119 |
Linearity
To assess the linearity of the assay, samples spiked with high concentrations of mouse BLC/BCA-1 in each matrix were diluted with Calibrator Diluent and then assayed.
Scientific Data Images for Mouse CXCL13/BLC/BCA-1 Quantikine ELISA Kit
Mouse CXCL13/BLC/BCA-1 ELISA Standard Curve
Preparation and Storage
Shipping
Stability & Storage
Background: CXCL13/BLC/BCA-1
Alternate Names
Gene Symbol
Additional CXCL13/BLC/BCA-1 Products
Product Documents for Mouse CXCL13/BLC/BCA-1 Quantikine ELISA Kit
Product Specific Notices for Mouse CXCL13/BLC/BCA-1 Quantikine ELISA Kit
For research use only
⚠ WARNING: This product can expose you to chemicals including N,N-Dimethylforamide, which is known to the State of California to cause cancer. For more information, go to www.P65Warnings.ca.gov.