Protein A ELISA Kit
R&D Systems, part of Bio-Techne | Catalog # BPPA00
Key Product Details
Assay Length
Sample Type & Volume Required Per Well
Sensitivity
Assay Range
Product Summary for Protein A ELISA Kit
Product Specifications
Measurement
Detection Method
Conjugate
Reactivity
Cross-reactivity
Interference
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in twenty separate assays to assess inter-assay precision. Assays were performed by at least three technicians using two lots of components.
Biosimilar Therapeutic Antibody
Intra-Assay Precision | Inter-Assay Precision | |||||
---|---|---|---|---|---|---|
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 20 | 20 | 20 |
Mean (pg/mL) | 157 | 507 | 1519 | 148 | 458 | 1442 |
Standard Deviation | 4.90 | 13.6 | 42.3 | 12.2 | 28.9 | 86.8 |
CV% | 3.1 | 2.7 | 2.8 | 8.3 | 6.3 | 6.0 |
Recovery for Protein A ELISA Kit
The recovery of Protein A spiked into biosimilar therapeutic antibody samples to levels throughout the range of the assay was evaluated. Each user should qualify their sample matrices for accurate recovery. For sample matrices other than PBS, further qual
Sample Type | Average % Recovery | Range % |
---|---|---|
Biosimilar Therapeutic Antibody (n=4) | 101 | 88-113 |
Linearity
To assess the linearity of the assay, biosimilar therapeutic antibody samples spiked with high concentrations of Protein A were serially diluted with calibrator diluent to produce samples with values within the dynamic range of the assay.
Scientific Data Images for Protein A ELISA Kit
Protein A Standard Curve
Preparation and Storage
Shipping
Stability & Storage
Background: Protein A
Additional Protein A Products
Product Documents for Protein A ELISA Kit
Product Specific Notices for Protein A ELISA Kit