Mouse IL-10 ELISA Kit - Quantikine Best Seller
R&D Systems, part of Bio-Techne | Catalog # M1000B
Key Product Details
Assay Length
Sample Type & Volume Required Per Well
Sensitivity
Assay Range
Product Summary for Mouse IL-10 Quantikine ELISA Kit
Product Specifications
Measurement
Detection Method
Conjugate
Reactivity
Specificity
Cross-reactivity
Interference
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision.
Inter-Assay Precision (Precision between assays) Three samples of a known concentration were tested in separate assays to assess inter-assay precision over two lots of kit components. Assays were performed by at least three technicians.
Cell Culture Supernates, EDTA Plasma, Heparin Plasma, Serum
Intra-Assay Precision | Inter-Assay Precision | |||||
---|---|---|---|---|---|---|
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 20 | 20 | 20 |
Mean (pg/mL) | 76.3 | 123 | 815 | 71.4 | 115 | 809 |
Standard Deviation | 2.56 | 2.85 | 17.0 | 5.39 | 7.64 | 55.5 |
CV% | 3.4 | 2.3 | 2.1 | 7.5 | 6.6 | 6.9 |
Recovery for Mouse IL-10 Quantikine ELISA Kit
The recovery of IL-10 spiked to levels throughout the range of the assay was evaluated.
Sample Type | Average % Recovery | Range % |
---|---|---|
Cell Culture Samples (n=4) | 99 | 95-103 |
EDTA Plasma (n=4) | 91 | 86-99 |
Heparin Plasma (n=4) | 89 | 82-98 |
Serum (n=4) | 89 | 85-96 |
Linearity
To assess the linearity of the assay, samples containing and/or spiked with various concentrations of mouse IL-10 in each matrix were diluted with Calibrator Diluent and then assayed.
Scientific Data Images for Mouse IL-10 Quantikine ELISA Kit
Mouse IL-10 ELISA Standard Curve
Preparation and Storage
Shipping
Stability & Storage
Background: IL-10
Long Name
Alternate Names
Entrez Gene IDs
Gene Symbol
Additional IL-10 Products
Product Documents for Mouse IL-10 Quantikine ELISA Kit
Product Specific Notices for Mouse IL-10 Quantikine ELISA Kit
For research use only
⚠ WARNING: This product can expose you to chemicals including N,N-Dimethylforamide, which is known to the State of California to cause cancer. For more information, go to www.P65Warnings.ca.gov.